49
Parameter Setting
TC Process Medium
Note: HOLD mode active
Menu Display
TC process medium
TC process medium
You can choose from:
• Linear (input of TC coefficient)
• EN 27888
• Ultrapure water
(add. function SW 3400-008)
Module CONDI 3400-051 (Administrator)
Return
TC process medium
Sensor data
Cal preset values
Concentration
Input filter
Block
TC process medium (Administrator)
20.1 °C
Abort
TC
Off
Linear
1.225 mS/cm
Ultrapure water
EN 27888
OK
TC process medium (Administrator)
20.1 °C
Abort
1.225 mS/cm
OK
TC correction
1.223
20.1
Outp I1 5.70 mA
Favorites menu
TC
EN 27888: Natural waters (0 ... 35 °C)
Ultrapure water: traces of impurity
Reference temperature
Impurity
NaOH
NH3
HCl
NaCl
When the TC correction for process
medium is switched on, “TC” appears
in the display in measuring mode.
When you have selected “Ultrapure
water”, you must specify the type of
impurity:
NaOH
Alkaline ultrapure water
NaCl
Neutral ultrapure water, for
conductivity measurement in water
processing behind gravel bed filter
HCl
Acidic ultrapure water, for
conductivity measurement behind
cation filter
NH3
Ammoniacal ultrapure water
20.1 °C
1.225 mS/cm
Messages