Menu Display
Manual entry of cal solution
Select calibration menu
Select “Module CONDI”
Select calibration method
“Manual entry of cal solution”,
confirm with enter.
Enter process temperature, if manual
temperature adjustment has been
selected.
Immerse sensor in cal solution.
Start calibration with softkey or
enter.
Enter conductivity.
End calibration
using softkey (”End”).
Calibration is running.
The display shows:
• Calibration temperature
• Response time
Module CONDI 3400-051
Return
25.0 °C
1.225 mS/cm
Product calibration
Data entry - premeasured sensor
Automatic with standard cal solution
Manual entry of cal solution
Manual entry
Dip sensor in cal solution
then `Start` calibration
24.9 °C
1.225 mS/cm
Start
Return
Measured cal temp
+24.9°C
Manual entry
Calibration running
Determine pair of cond/°C values
24.9 °C
1.225 mS/cm
Calibration temperature
+24.9°C
Response time
0001s
Manual entry
25.0 °C
1.225 mS/cm
End
Repeat
Conductivity
Enter calibration solution
Measured cal temp
for correct temperature!
+025.0 °C
0251 mS/cm
Sensor replacement
Calibration data record
Calibration
25.6 °C
1.225 mS/cm
Cal mode
03/31/10 11:37
Cell factor
Manual input
Adjust
End
Adjustment
Press “Adjust” to take over the values
determined during calibration for
calculating the measured variables.
0.988 S/cm
Zero correction
Zero
+000.0 µS/cm