Menu Display
Automatic calibration
Select calibration menu
Select “Module CONDI”
Select calibration method “Automatic
with standard cal solution”,
confirm with enter.
Display of selected calibration
solution.
Enter process temperature, if manual
temperature adjustment has been
selected.
Dip sensor in calibration solution.
Start calibration with softkey or
enter.
Calibration is running.
The display shows:
• Calibration temperature
• Solution table value
(conductivity in dependence on
process temperature)
• Response time
Rinse sensor and replace it in the
process, end calibration with softkey
or enter
Module CONDI 3400-051
Return
25.0 °C
1.225 mS/cm
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Automatic with standard cal solution
Automatic
Dip sensor in cal solution
25.0 °C
1.225 mS/cm
TC automatically considered
then “Start” calibration.
Start
NaCl saturated
Return
Cal solution
Automatic
Calibration running
25.6 °C
1.225 mS/cm
Correction of cell factor
Calibration temp
+025.0 °C
Solution table value
Response time
0003s
Enter cal temp
+025.0 °C
0.020 mS/cm
Repeat
End
Calibration data record
Calibration
25.6 °C
1.225 mS/cm
Cal mode
03/31/10 11:37
Cell factor
Automatic
Adjust
End
Adjustment
Press “Adjust” to take over the values
determined during calibration for cal-
culating the measured variables.
0.988 S/cm
Sensor replacement
Zero correction
Zero
+000.0 µS/cm