22
Programming the HPLC Pump K–1001
01
07
100.0
|200| 0 0 0 0 0 0 0 0
OFF
000.00
100 0
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
A
01
07
100.0
|200|
0 0 0 0 0 0 0 0
OFF
000.00
90
10
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
B
01
07
- - -.-
|200| -- -- -- -- -- -- -- --
OFF
***.*
- -
- -
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
C
01
07
_ _ _._
|200|
_ _ _ _ _ _ _ _
OFF
020.0
10
90
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
D
Fig. 16
Display of sinle steps, when creating a program
(see „Example for creating a program“on page 24)
SOP 14 Execution of Programs
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X.
1. Load the program you want to execute by entering the program number in
the LOAD PROG-field. Confirm with any arrow key.
2. Press START key to start the pump. The display changes to
ON
and the
pump begins continuous operation with the parameters specified in the first
program line.
3. Position cursor on field
ON
4. Start the program by pressing the numeric key 1. The display changes to
RUN
.
5. During execution of programs you can press key 0 to freeze the current
status of the pump. The display changes to
HOLD
and the field TIME is
blinking.
6. If you press 1 you can activate program execution again or you can press
STOP
to abort execution.
During program execution the current solvent composition can be
observed in fields % A and % B. At the end of the program the pump
returns continuous pumping with the parameters of the first line, ON is
displayed.