12
Control of the HPLC Pump K–1001
1. Place the cursor in a field in one of the four corners of the main screen,
e.g. on
OFF
in the lower left corner of the main menu.
2. Depending on the selected corner, press the corresponding left or right
arrow key longer than 2 seconds. In the example above selection of key
½
will activate the VIEW Menu on the display.
3. Select the rhombus
in the lower left corner of the display if you want to
access additional pages of the selected menu. You can access all menu
pages by scrolling using
À
or
¿
arrow keys.
4. Select the rhombus
and use
¾
or
½
to return to the main screen.
Links:
21
___
___
___
___
(21 . . 29)
___
___
___
___
Load
Edit Prog
Flow
Pressure
Events
Prog
Clear 0
[ml/min]
[0.1 Mpa]
off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
Raute
Fig. 7
Example of a multi-page menu: VIEW menu
Menus without rhombus are single page menus.
SOP 2
Entering Data
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X. Use this
SOP for entering parameters and program steps in your HPLC Pump K–1001.
1. Place the cursor at the desired position of the main screen.
2. Press a numeric key to start data input. The appearance of the cursor
changes to a underline character, see Fig. 5 on page 11.
3. Press numeric keys for entering desired numeric values. Wrong entries
can simply be overwritten by continuing the entry.
4. Confirm value by pressing any of the arrow keys. This finishes data entry
mode and moves the cursor to the next field according to the arrow key
direction.
5. You can modify and delete data according to SOP 15 „Modifying and
Deleting Programs“ on page 23.
Using the Arrow keys in the SETUP Menu
Fig. 8 „Structure of the SETUP Menu“ on page 13 gives a complete overview of
the Setup Menu.
You can use
¾
for accessing menus.
You can use
¿
and
À
for scrolling through the list of parameters of any
individual menu item.
Using the SETUP Menu
The SETUP menu of the HPLC Pump K–1001 enables access to the following
features and functions:
GRADIENT MODE:
The HPLC Pump K–1001 is capable of isocratic (none),
high pressure gradient (
HPG
) and low pressure gradient (
LPG
) operation.
You can choose the mode of operation: