22
Software Control of the PDA Detector 2800
Fig. 24
Detector Setup
In the Detector Setup the following settings can be entered:
•
Name (default: Det1)
•
Channels: Up to four separate channels can be selected for data
acquisition (default: one channel)
•
Process: Enables the measured detector signal to be evaluated
•
Averaging: Enables/disables averaging (default: enabled)
•
Integration time (ms): As a rule integration time should not
exceed 100 ms. Otherwise there may be a problem with the light
intensity.
•
Begin/End scan at: Can be chosen within the detector’s individual
scan range (default: 200 nm – 600 nm)
•
Analog Output Scale: Which absorption (in AU) should equal 10
V on the analog output(s) can be selected here
•
Data: Select whether Absorption, Emission, Transmission or
ADC Counts should be recorded. (default: Absorption)
•
Shutter Control: Shutter(s) can be enabled or disabled. (default:
disabled)
By clicking on OK all settings are saved and the Detector Setup menu is
closed.