Software Control of the PDA Detector 2800
17
Fig. 16
Detector Configuration
From the Detector Configuration menu it is possible to change settings for:
•
Name (default: S-2800)
•
Interface: Ethernet or PCI (for S-2800 only Ethernet can be used)
•
Y-Axis Units: AU, mAU or µAU (default: mAU)
•
Serial Number: The serial number of the detector can be entered
(default: 0)
•
IP Address: Here you must enter the Instrument’s IP address
(default: 172.16.5.241)
•
Flow Cell: The type of flow cell installed can be entered here
(default: blank)
•
D2-Lamp: This option is always activated
•
Halogen Lamp: Please activate this option if a halogen lamp is
installed
•
Shutter Control: if this box is checked, shutter control is enabled
(default: disabled)
By clicking on OK the changes are saved and the Detector Configuration
menu is closed.
Fig. 17
Configuration Options
If you wish to work in PDA mode (spectra acquisition enabled) the PDA
option must be activated from the Options menu.
(PDA option can only be activated if a PDA license has been previously
installed!)