Technical Data
39
P 2.1S/P 4.1S user manual V6870, version 3.1
Technical Data
Solvent delivery
Pump type
Dual-piston pump with main and
auxiliary piston
Flow rate range
10 ml pump head:
0.001 - 10 ml/min
50 ml pump head:
0.01 - 50 ml/min
Maximum pressure
10 ml pump head:
40 MPa to 10 ml/min
Imax = 70
50 ml pump head:
15 MPa to 50 ml/min
Imax = 80
Flow rate accuracy
± 1%, measured at 5 - 50% of
flow range using ethanol/water
10:90
For pumps without a pressure
sensor dependent on pressure
Flow rate precision
Relative standard deviation RSD:
< 0.5 % (1 ml/min)
Gradients
Isocratic HPLC pump
Expandable to a binary high pres-
sure gradient system (controlled
by software)
System protection
Pump with pressure sensor:
P
min
and P
max
adjustable
I
min
and I
max
adjustable
Pump with pressure sensor:
I
min
and I
max
adjustable
Communication
Operation
LAN
RS-232
Analog
Buttons on the device
Analog inputs
0 – 10 V
Supply frequency
50-60 Hz
Active power con-
sumption
maximum 40 W
Protection
IP-20