30
Operation
P 2.1S/P 4.1S user manual V6870, version 3.1
Starting with a Short-Circut
Connection for the short circuit (or TTL-low) for interrupting and continuing the
operation of the pump.
The operation of the pump is dependent on the
STARTLEVEL setting:
STARTLEVEL 1
(standard setting RS-232, LAN)
: The pump does not operate
during the time of short-circuit.
STARTLEVEL 0
(standard setting Analog)
: The pump operates during the time
of short-circuit.
Enter
STARTLEVEL:1
in the terminal program,
to stop the pump during short-cir-
cuit
.
Prerequisite:
The pump is connected to the power supply.
Note:
During the interruption, the horizontal arrow remains in the display,
because the pump is still in operation status.
Process:
1. Connect the pump to a suitable terminal program.
2. Connect the pump using a LAN or RS-232.
3. Enter
STARTLEVEL:1
.
Starting Directly after Connecting to Power Supply
By default the pump is stopped and started using the start/stop button. The
STARTMODE setting allows you to start the device right after connecting it to
power supply.
STARTMODE 0
(default setting RS-232, LAN)
: The pump does not start operat-
ing right after being connected to power supply.
STARTMODE 1
(default setting Analog)
: The pump starts operating right after
being connected to power supply.
Enter in the terminal program using RS-232
STARTMODE:1
, so that the pump is
in operation directly after being connected to the power supply.
Prerequisite:
The pump is connected to the power supply.
Process:
1. Connect the pump to a suitable terminal program.
Fig. 23
Status display
Legend
1
Pressure in 0,1 MPa
2
Flow rate in ml/min
1
2