Klinger CMF
Instruction Manual
Page 41
5.2.5.3.28.3.1 When the fluid direction is set to the FWD, reverse or absolute value, Qmax
and Qmin should be
≥
0, and Qmax > Qmin.
5.2.5.3.28.3.2 When the fluid direction is set “2 WAY”, Qmax > 0, Qmin < 0, See table 5-2.
Table 5
-2 Mass flowmeter flow direction
Mass flowmeter flow direction
Flow
direction
setting
condition
display
Qmax
Qmin
Output
(mA)
Cumulative
flow
FWD
Flow matches direction
of forward flow
indicator
Flow
0~99999
0
4 to 20
Totalizer
increased
Flow opposite forward
flow indicator
0
4
No change
REV
Flow matches direction
of forward flow
indicator
0
0~99999
0
4
No change
Flow opposite forward
flow indicator
Flow
4~20
Totalizer
increased
ABS
Flow matches direction
of forward flow
indicator
Flow
0~99999
0
4~20
Totalizer
increased
Flow opposite forward
flow indicator
Flow
4~20
Totalizer
increased
2 WAY
Flow matches direction
of forward flow
indicator
﹢
Flow
0~99999
-99999~0
4~20
Totalizer
increased
Flow opposite forward
flow indicator
-
Flow
Totalizer
decreased
Note:1:Qmax-Qmin≥20% flow range (see PDS);2:Qmin =0~Qmax
(
non bi-directional flow setting
)
5.2.5.3.29 COMM
5.2.5.3.29.1 Set flow meter communication mode.
5.2.5.3.29.2 Option: RS485 (Modbus RTU), HART.
5.2.5.3.30 CONCTRN
5.2.5.3.30.1 Input raw data for calculating the concentration.
5.2.5.3.30.2 CAL PTS: input points,0~5
5.2.5.3.30.3 T/DA/DB: input Temperature
,
Density(A) & Density(B) for each point.
5.2.5.3.30.4 Temperature input range:-50
℃
~300
℃
5.2.5.3.30.5 Density input range:0g/cm3~3 g/cm3
5.2.5.3.31 RESET MEMORY
Restore the configuration to the factory default.