Configure the measured variable & unit (see table 5-1) displayed in secondary display .
Table 5-1 Display variable and corresponding unit
Display variable
Optional unit
Default unit
Mass flow (Fm)
g/s
g/min
g/h
kg/s
kg/min
kg/h
kg/h
kg/day
t/s
t/min
t/h
t/day
lb/s
lb/min
lb/h
lb/day
Volume flow (Fv)
mL/s
mL/min
mL/h
L/s
L/min
L/h
L/h
L/day
m
3
/s
m
3
/min
m
3
/h
m
3
/day
USGPS
USGPM
USGPH
USGPD
Total mass (Σm)
g
kg
t
lb
kg
Total volume (Σv)
mL
L
m
3
USGAL
L
Density (D)
g/cm
3
g/L
kg/L
kg/m
3
t/m
3
g/cm
3
lb/gal
Temperature (T)
℃
℉
℃
5.2.5.3.5 DIGITS
Set the number of decimals to display on the primary and secondary display, the range:
0-3.
5.2.5.3.6 UPDATE T
Set the damping time. The output is the average value measured across the interval.
5.2.5.3.6.1 Refresh damping setting range: 0s-20s.
5.2.5.3.6.2 If the filter coefficient is less than ATH, the flow is stable, and at this time, the
average output of the refresh time is adopted.
5.2.5.3.6.3 If the filter coefficient is greater than ATH, that the flow in the dynamic process
and at this time the refresh time is invalid, the output value is the direct
measurement value.
5.2.5.3.6.4 If the site noise is high & the filter coefficient is keeping higher than ATH, that
means refresh time keeping invalid. Following the ATH can be set higher, allowing
the damping output Under the condition of a higher filter coefficient.
5.2.5.3.7 MASS CUTOFF
5.2.5.3.7.1 Set the mass low flow cutoff ratio. Setting range 0%~50% (“M.FLOW” in "INFO").
5.2.5.3.7.2 The default value is 1%
5.2.5.3.7.3 When “mass flow”< (“MASS CUTOFF”x “M.FLOW”), the output mass flow is driven
to “zero”.
5.2.5.3.8 VOL CUTOFF
5.2.5.3.8.1 Set the volumetric low flow cutoff ratio. Setting range 0%-50% (“V.FLOW” in
"INFO").
5.2.5.3.8.2 The default value is 1%.
5.2.5.3.8.3 When “volume flow”< (“VOL CUTOFF”x “V.FLOW”), the output volume flow is
driven to “zero”.
5.2.5.3.9 DEN CUTOFF
Set the low density cutoff value (0g/cm
3
to 3g/cm
3)
. The output is driven to “0” while
the measured densities below the value.
5.2.5.3.10 FIXED DENS
5.2.5.3.10.1 When the input value is “0”, the displayed density is the measured value, “volume