Klinger CMF
Instruction Manual
Page 20
26
2.2.3 Transmitter Installation
2.2.3.1 Installation
2.2.3.1.1 Integrated installation
2.2.3.1.1.1 For the installation of an integrated mass flow meter, there is no special
transmitter installation guide.
2.2.3.1.1.2 To enhance visibility the transmitter can be rotated up to 360 degrees in 90 degree
increments. (Please confirm the original state before rotating) To do so:
Loosen and remove the bolts connecting the transmitter and the flange it is
mounted to.
Gently rotate the transmitter to the desired angle.
Screw the transmitter to its flange.
2.2.3.1.1.3 To enhance visibility the LCD can be rotated up to 270 degrees according in 90
degree increments mode. (Please confirm the original state before rotating) To do
so:
Remove the front cover from the transmitter.
Loosen and remove the screws securing the LCD to the transmitter.
Check the LCD connections for pre-existing twist.
Gently rotate the LCD to the desired angle ensuring that the connections do
not become over twisted.
Re-install and tighten the LCD screws and the front cover.
2.2.3.1.2 Remote installation
2.2.3.1.2.1 Transmitter can be installed in any desired location.
2.2.3.1.2.2 Cable length: 5 m -100 m.
2.2.3.2 Instructions and precautions
2.2.3.2.1 Please do not place the transmitter conduits up along.
2.2.3.2.2 Ensure that the cover of transmitter is fully tightened.
2.2.3.2.3 Ensure that the O-ring is firmly seated and in good condition.