A-8
PCZ1000A
The error occurs.
What to Check
Result
Possible Causes and Countermeasures
Page
Is the "xx xxx EPPDATA
ERR" displayed on the
panel? ("x" may varies
depends on the contents of
the error)
Yes
The calibration data is changed to the abnormal value due to the
internal memory is deteriorated, arrange to have the PCZ1000A
repaired.
No
Abnormal condition has been occurred in the internal circuit.
Immediately stop using the PCZ1000A and arrange for it to be
repaired.
-