3-32
PCZ1000A
4
Press the A key to display " GAIN XXXXX".
XXXXX is the RMS gain. The display switches between "IRS OFFS XXXXX" and
"IRS GAIN XXXXX" each time the A key is pressed.
5
Use the rotary knob to adjust the gain value (variable range: 31 000 to
39 000) on the front panel display to the value on the externally
connected AC ammeter (RMS meter).
As the offset and gain are mutually dependent, it is recommended that you adjust each
in turn.
6
After calibrating the offset and gain, press the MEM key to display
"SAVE?".
If you want to stop calibration before finishing, press the MEM key to display
"SAVE?" and then press the ESC key.
7
Press the ENTER key.
The calibration data is written to internal memory in the PCZ1000A.
• If the calibration can not be performed for the offset value and the gain value
within the variable range, please confirm whether the POWER switch of the
slave unit is turned in the "off" position. In case the POWER switch of the slave
unit is not in the "off" position, contact your Kikusui distributor/agent.
■
Calibrating the offset and gain for the total current (peak)
1
Press the R SET key to display "IPS OFFS XXXXX".
XXXXX is the peak offset value.
2
Press the LOAD key to change the PCZ1000A to the load-on state.
3
Use the rotary knob to set the offset value on the front panel display to
the value on the externally connected AC ammeter (peak meter).
4
Press the A key to display "IPS GAIN XXXXX".
XXXXX is the RMS gain. The display switches between "IPS OFFS XXXXX" and
"IPS GAIN XXXXX" each time the A key is pressed.
5
Use the rotary knob to adjust the gain value on the front panel display to
the value on the externally connected AC ammeter (peak meter).
As the offset and gain are mutually dependent, it is recommended that you adjust each
in turn.
6
After calibrating the offset and gain, press the MEM key to display
"SAVE?".
If you want to stop calibration before finishing, press the MEM key to display
"SAVE?" and then press the ESC key.
7
Press the ENTER key.
The calibration data is written to internal memory in the PCZ1000A. This completes
calibration.
Summary of Contents for PCZ1000A
Page 4: ...ii PCZ1000A This page is intentionally blank ...
Page 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Page 35: ...Operation This chapter describes each mode and function of the operation ...
Page 82: ...5 4 PCZ1000A This page is intentionally blank ...
Page 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Page 88: ...6 6 PCZ1000A This page is intentionally blank ...
Page 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Page 99: ......