3-34
PCZ1000A
3
Press SHIFT + I SET (CONFIG) or the ESC key to exit CONFIG
setting.
• When setting a slave unit back to "NORMAL", perform the Step 1 to Step 3
operations within 30 seconds. An external alarm 2 occurs if this procedure is not
performed within 30 seconds.
3.8
Factory Default Settings
Hold down the MEM key as you turn on the POWER switch and do not release until
after about 8 seconds. First, the model name and firmware version are displayed,
then the screen shown in Fig.3-12 appears for approximately 2 seconds and the
instrument is set back to its factory default settings.
Fig.3-12 Completion Screen When Restoring Factory Default Settings
The table below lists the factory default settings. The A, B, and C preset memories
are also restored to their factory default settings.
Table3-3 Factory Default Settings
Setting
Setting Value (Panel Setting)
Current limit
0 A
Resistance limit
1 000.0
Ω
Power limit
1 050 W
Crest factor
1.4
Operation mode
CC mode
CR mode range
H
Crest factor function on/off
OFF
ABC preset memory
Above settings for each mode
Load on/off
Load-off
CONFIG
1. Norm/M/S
NORMAL
2. Para/Track
PARALLEL
3. Parallel Ope.
2
4. Calibration
Push [I SET] (fixed)
PCZ1000A
**** RESET ****
Summary of Contents for PCZ1000A
Page 4: ...ii PCZ1000A This page is intentionally blank ...
Page 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Page 35: ...Operation This chapter describes each mode and function of the operation ...
Page 82: ...5 4 PCZ1000A This page is intentionally blank ...
Page 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Page 88: ...6 6 PCZ1000A This page is intentionally blank ...
Page 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Page 99: ......