Troubleshooting Procedure
4
Keysight X8711A IoT Device Functional Test Solution User’s Guide
48
Test Plan Execution Error Messages
Unable to load type ‘<Test Step>’
Depending on your choice of TAP software, you would observe the following
message on your GUI or Command Prompt window.
Unable to load type ‘<Test Step>>’...
You may have installed incompatible plugins into the same TAP environment. Here
are a few tips that you can try:
– Backup folder
%TAP_PATH%\IoTDeviceTest\TapFiles
.
– Remove the entire
%TAP_PATH%\IoTDeviceTest
directory.
– Reinstall the measurement suite plugins.
– Restore the TapFiles folder from backup.
– Verify Bench profile settings.
Could Not Create ‘<Test Step>’ Step
You will see the following message when you do not have the measurement suite
plugins license.
You will need to purchase the following license that are offered in the X8711A
– KS83301A or KS83301B:
Bluetooth
®
LE related test steps
– KS83302A or KS83301B: WLAN related test steps
Once you have purchased the license(s), you will receive an entitlement certificate
via email. Follow the instructions in the email to redeem the license.