Test Sequence Application
7
Keysight U8903B User’s Guide
381
Signal generation
The receiver sensitivity signal generation settings menu page is displayed as shown in
.
Figure 7-54
Receiver Sensitivity > Signal Generation settings menu page
Table 7-52
Receiver Sensitivity > Signal Generation settings menu description
Menu
Description
Instrument
Press the
Instrument
softkey to select the signal generator model.
– Keysight E4438C
– Other
GPIB address
Press the
GPIB address
softkey to set the GPIB address for the connected signal generator.
Start Power
Press the
Start Power
softkey to set the sweep start RF power.
Stop Power
Press the
Stop Power
softkey to set the sweep stop RF power.
Step Size
Press the
Step Size
softkey to set the step size.
Dwell Time
Press the
Dwell Time
softkey to set the delay between each measured SINAD in seconds.
Init Instrument
Press the
Init Instrument
softkey to enable or disable the U8903B to send SCPI commands to initialize the connected
signal generator.
FM Frequency
Press the
FM Frequency
softkey to set the RF signal output frequency. This setting is only available if
Instrument
is set to
Keysight E4438C and the
Init Instrument
is enabled.
FM Deviation
Press the
FM Deviation
softkey to set the RF signal frequency modulation deviation. This setting is only available if
Instrument
is set to Keysight E4438C and the
Init Instrument
is enabled.
FM Rate
Press the
FM Rate
softkey to set the RF signal internal frequency modulation rate. This setting is only available if
Instrument
is set to Keysight E4438C and the
Init Instrument
is enabled.
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...