Audio Analyzer Measurement Functions
240
Keysight U8903B User’s Guide
Measurement Configuration (Digital Analyzer)
The measurement configuration for digital analyzer is displayed as shown in
Figure 4-57
Digital Analyzer > Meas Config menu page
Table 4-28
Digital Analyzer > Meas Config menu description
Menu
Description
Coupling
Press the
Coupling
softkey to select the coupling type.
– DC
DC coupling allows both AC and DC input signals to pass through to the digital analyzer and to be measured down to
0 Hz. This setting should be selected when making DC voltage measurements.
– AC
AC coupling blocks the DC component of the input signal. This setting should be selected when you need to measure
only the AC component of a signal. For example, when you are making rms or peak-to-peak voltage measurement.
Sample Size
Press the
Sample Size
softkey to select the sample size. The sample size is the number of data points acquired for the
measurement. The higher the sample size, the higher the accuracy of the measurement result.
– 2048
– 4096
– 8192
– 16384
– 32768
– 65536
– 131072
– 262144
– 524288
– 1M
– 2M
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...