Audio Analyzer Measurement Functions
4
Keysight U8903B User’s Guide
199
Format
Press the
Format
softkey to select the format of the returned measurement reading.
– Off
Selecting Off will return the measurement reading without any formatting (raw value).
– Logarithmic
Selecting Logarithmic will return the measurement result by using the following formula:
The result is returned in unit dBr.
– Linear
Selecting Linear will return the measurement result by dividing the measured level by the reference level. The result is
returned in unit x.
– Delta
Selecting Delta will return the meausrement result by deducting the reference level from the measured level. The result
is returned in unit
Δ
V. This is only available for the analog analyzer.
Set to 0dB
This is only available for analog analyzer.
Press the
Set to 0dB
softkey to store the measured level as the reference level, and set the measurement reading format to
Logarithmic.
Set result as ref. from
Press the
Set result as ref. from
softkey to store the measurement result from the selected channel as the reference level.
Ref. Level
Press the
Ref. Level
softkey to set the reference level value. There is only one reference level for each channel.
Ref. Imp
This is only available for the analog analyzer.
Press the
Ref. Imp
softkey to set the reference impedance value. This setting is only available if the unit is set to W and
dBm.
Volt/FS
This is only available for the digital and
Bluetooth
analyzer.
Press the
Volt/FS
softkey to set the volts/FS reference value.
Precision
This is only available for the digital analyzer.
Press the
Precision
softkey to enable or disable the precision mode.
Cal SPL
Press the
Cal SPL
softkey to set the calibrator level value. This setting is only available when the unit is set to dBSPL.
Freq Lock
Press the
Freq Lock
softkey to select the searching method for the fundamental frequency.
– Auto
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically.
– Gen Lock
Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency
value of the respective generator channel. You can select the generator channel by setting the
Src Channel
in the
Meas.
Config.
menu page. This Gen Lock method is only useful if you use the internal audio generator.
– Custom
If you have a known input signal, you can define the fundamental frequency value by setting the searching method to
Custom and setting the frequency value in
Fund Freq
.
Fund Freq
Press the
Fund Freq
softkey to set the fundamental frequency value. This setting is only available when
Freq Lock
is set to
Custom.
Analog Notch
This is only available for analog analyzer.
Press the
Analog Notch
softkey to enable or disable the analog notch filter. This filter is used to emulate the analog notch
filter used in HP8903B.
Table 4-10
Analog/Digital/
Bluetooth
Analyzer > Functions > THD+N Level menu description (continued)
Menu
Description
20
10 Measured rms voltage
Reference level
-------------------------------------------
⎝
⎠
⎛
⎞
log
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...