![Keysight U8903B User Manual Download Page 196](http://html.mh-extra.com/html/keysight/u8903b/u8903b_user-manual_1961457196.webp)
4
Audio Analyzer Measurement Functions
196
Keysight U8903B User’s Guide
THD+N level is defined as the square root of the sum of the squares of all the signal
harmonics components and noise amplitude, and it is expressed in Vrms. The THD+N level
can be computed as follows.
Table 4-9
Analog/Digital/
Bluetooth
Analyzer > Functions > THD+N Ratio menu description
Menu
Description
Unit
Press the
Unit
softkey to select the measurement unit.
– dB
–
Δ
dB
– %
– x
Format
Press the
Format
softkey to select the format of the returned measurement reading.
– Off
Selecting Off will return the measurement reading without any formatting (raw value).
– Linear
Selecting Linear will return the measurement result by dividing the measured ratio by the reference ratio. The result is
returned in unit x.
– Delta
Selecting Delta will return the measurement result by deducting the reference ratio from the measured ratio. The result
will be returned in unit
Δ
dB.
Set to 0dB
This is only available for analog analyzer.
Press the
Set to 0dB
softkey to store the measured ratio as the reference ratio, and set the measurement reading format to
Delta.
Set result as ref. from
Press the
Set result as ref. from
softkey to store the measurement result from the selected channel as the reference ratio.
Ref. Ratio
Press the
Ref. Ratio
softkey to set the reference ratio value.
Precision
This is only available for digital analyzer.
Press the
Precision
softkey to enable or disable the precision mode.
Freq Lock
Press the
Freq Lock
softkey to select the searching method for the fundamental frequency.
– Auto
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically.
– Gen Lock
Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency
value of the respective generator channel. You can select the generator channel by setting the
Src Channel
in the
Meas.
Config.
menu page. This Gen Lock method is only useful if you use the internal audio generator.
– Custom
If you have a known input signal, you can define the fundamental frequency value by setting the searching method to
Custom and setting the frequency value in
Fund Freq
.
Fund Freq
Press the
Fund Freq
softkey to set the fundamental frequency value. This setting is only available when
Freq Lock
is set to
Custom.
Analog Notch
This is only available for analog analyzer.
Press the
Analog Notch
softkey to enable or disable the analog notch filter. This filter is used to emulate the analog notch
filter used in HP8903B.
THD+N Level
rms value of noise and distortion
=
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer User s Guide ...
Page 16: ...16 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 32: ...32 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 278: ...5 Graph Analysis 278 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 290: ...6 Sweep Function 290 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...