50
Chapter 2
Test Descriptions
Frequency Accuracy
Step 6. Use Cells D1 to D201 as a look up reference and find the cell, which
closest matches the ½ Power Reading in cell D202.
The corresponding frequency in Column A is given as the center
frequency point. Enter the reading into the appropriate column in the
Test Record. Ensure that the measured value meets its published
specification.
Performing the frequency accuracy calculations manually
Step 1. Using the displayed P
hot
results sum the values and enter the summed
values on the Test Worksheet Table on page 45.
Example 2-2
1 GHz CW at the 4 MHz bandwidth
For the 1 GHz CW at the 4 MHz bandwidth each Phot reading shown in
the Figure 2-12 on page 51 is summed with the previous cell(s) until the
total power under the curve is equated in the lower right hand column.
Example sum reading 1 to reading 6 = 0.369.
Figure 2-11
Displayed Table Values
Summary of Contents for N8972A
Page 9: ...ix Declaration of Conformity N8972A ...
Page 10: ...x Declaration of Conformity N8973A ...
Page 11: ...xi Declaration of Conformity N8974A ...
Page 12: ...xii Declaration of Conformity N8975A ...
Page 19: ...Contents xix ...
Page 20: ...xx Contents ...
Page 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Page 124: ...104 Chapter3 Technical Specifications General Specifications ...
Page 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Page 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Page 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Page 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...