54
Chapter 2
Test Descriptions
Noise Source Supply Accuracy
Step 8. Set the noise source on (press the
System
key, then
More
,
More
and
Service
. Press
Noise Source On
). Press
Enter Password
then enter the
service password
-2010
. Press
Enter
then
Service
and
Noise Source On
).
Step 9. Enter the digital multimeter reading in the Test Record as the Noise
Source Supply On measurement.
Step 10. Press the
Preset
key to return to the default Noise Source Supply State.
Step 11. Verify that the measured values are within the published specifications.