56
Chapter 2
Test Descriptions
Noise Figure Range and Accuracy
Related Adjustment
None
Required Test Equipment
Table 2-12
Noise Figure Range and Accuracy test equipment
Equipment description
Critical specification for equipment
substitution
Recommended model
Synthesized Sweeper
Frequency Range: 50 MHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -54dBm to -65dBm
83620/30/40/50B
Option 001 and 008
Attenuator/Switch Driver
Compatible with 8494G/H Programmable step
attenuators
11713A
1 dB Step Attenuator
Attenuation Range: 0 to 11dB
Frequency Range: 50 MHz
Connectors: 3.5mm (F)
Calibrated at 50 MHz with accuracy of 0.010 dB
8494G/H Option 002
Adapter 1
N8972/3A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm (F)
1250-1744
1250-1749 or 83059B
Adapter
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
1250-1749 or 83059B
11901B
RF Cable (X 2)
50 MHz 3.5mm (M) to 3.5mm (M), 61cm (24 in.)
11500E
Summary of Contents for N8972A
Page 9: ...ix Declaration of Conformity N8972A ...
Page 10: ...x Declaration of Conformity N8973A ...
Page 11: ...xi Declaration of Conformity N8974A ...
Page 12: ...xii Declaration of Conformity N8975A ...
Page 19: ...Contents xix ...
Page 20: ...xx Contents ...
Page 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Page 124: ...104 Chapter3 Technical Specifications General Specifications ...
Page 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Page 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Page 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Page 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...