10 MHz Out Frequency Reference Accuracy
10 MHz Out Frequency Reference Accuracy
Test Description
The test applies to NFAs with either the standard 10 MHz frequency
reference or 10 MHz precision frequency reference (Option 1D5).
The test measures both the frequency reference accuracy and the ability
to set the timebase. The frequency reference accuracy is measured with
the NFA’s 10 MHz REF OUT connected to the frequency counter's
Channel A input. A cesium beam frequency standard, or some other
10 MHz ± “house standard” provides the frequency reference for the
frequency counter. The timebase to be set is measured by changing the
setting of the digital-to-analog converter (DAC) which controls the
frequency of the timebase. The frequency difference over the range of
DAC settings is calculated and compared to the specification.
Test Specification
Related Adjustment
10 MHz Frequency Reference Adjustment (see page 18).
Standard
Option 1D5
Ageing Rate
< ±2 ppm
/year
< ±1 ppm/year
Settability
< ±0.5 ppm
< ±0.01 ppm
Temperature Stability
< ±6 ppm
< ±0.01 ppm
Summary of Contents for N8972A
Page 9: ...ix Declaration of Conformity N8972A ...
Page 10: ...x Declaration of Conformity N8973A ...
Page 11: ...xi Declaration of Conformity N8974A ...
Page 12: ...xii Declaration of Conformity N8975A ...
Page 19: ...Contents xix ...
Page 20: ...xx Contents ...
Page 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Page 124: ...104 Chapter3 Technical Specifications General Specifications ...
Page 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Page 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Page 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Page 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...