Chapter 1
7
Preparing for Calibration and Performance Verification Test
Recommended Test Equipment
3.5mm Calibration Kit
Impedance: 50
Ω
Frequency Band: 3 GHz to 26.5 GHz
UUT: N8974A and N8975A
85052D
Synthesized Sweeper
Frequency Range: 10 MHz to 26.5 GHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -55 dBm
83620/30/40/50B
Option 001 or 008
Digital Multimeter
Input Resistance of 10 M
Ω
Accuracy of 10 mV on 100 V range
3458A
Attenuator/Switch Driver
Compatible with 8494G/H Programmable
step attenuators
11713A
1 dB Step Attenuator
Attenuation Range: 0 to 11 dB
Frequency Range: 50 MHz
Connectors: 3.5 mm (F)
Calibrated at 50 MHz with accuracy of
0.010 dB
8494G/H Option 2
Noise Source
Frequency Range:10 MHz to 6.7 GHz
Typical ENR: 4.5 - 6.5 dB
Connection: 3.5 mm(M) or Type N
346A Std or Option
001
Noise Source
Frequency Range: 10 MHz to 26.5 GHz
Typical ENR: 12.0 - 17.0 dB
Connection: 3.5 mm(M)
346C
Table 1-1
Recommended Test Equipment
Equipment description
Critical specification for equipment
substitution
Recommended
model
Summary of Contents for N8972A
Page 9: ...ix Declaration of Conformity N8972A ...
Page 10: ...x Declaration of Conformity N8973A ...
Page 11: ...xi Declaration of Conformity N8974A ...
Page 12: ...xii Declaration of Conformity N8975A ...
Page 19: ...Contents xix ...
Page 20: ...xx Contents ...
Page 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Page 124: ...104 Chapter3 Technical Specifications General Specifications ...
Page 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Page 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Page 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Page 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...