42 Installation Note N5247-90106
Installation Procedure for the Upgrade
-
Figure 26
Location of W145 Cable Securing Clamp
— W116 (N5247-20080) A45 port 2 bias tee to A36 port 2 coupler
— W121 (N5247-20024) A48 port 4 receiver attenuator to port 4 RCVR D IN
— W111 (N5247-20021) A44 port 4 bias tee to port 4 CPLR THRU
— W42 (N5247-20026) A35 port 4 coupler to port 4 CPLR ARM
— W110 (N5247-20025) A40 port 4 source attenuator to port 4 SOURCE OUT
— W41 (N5247-20069) A31 port 4 reference coupler to REF 4 SOURCE OUT
— W55 (N5247-20067) A28 mixer brick (R4) to REF 4 RCVR R4 IN
, install clamp part number 1400-1334 to secure
W41, and install cable tie part number 1400-0249 to secure W55.