1
Specifications and Characteristics
32
Keysight N1913/1914A Service Guide
Keysight E9301H
<±1.5 nW
<7 nW
Keysight N8481A (exclude Option CFT)
[d]
<±3 nW
<80 nW
Keysight N8482A (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8485A (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8486A R (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8486A Q (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8487A (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8488A (exclude Option CFT)
<±3 nW
<80 nW
Keysight N8481B (exclude Option CFT)
<±3 µW
<80 µW
Keysight N8482B (exclude Option CFT)
<±3 µW
<80 µW
Keysight N8481H (exclude Option CFT)
<±0.3 µW
<8 µW
Keysight N8482H (exclude Option CFT)
<±0.3 µW
<8 µW
Keysight N8481A with Option CFT
<±7 nW
<114 nW
Keysight N8482A with Option CFT
<±7 nW
<114 nW
Keysight N8485A with Option CFT
<±7 nW
<114 nW
Keysight N8486A R with Option CFT
<±7 nW
<114 nW
Keysight N8486A Q with Option CFT
<±7 nW
<114 nW
Keysight N8487A with Option CFT
<±7 nW
<114 nW
Keysight N8481B with Option CFT
<±7 µW
<114 µW
Keysight N8482B with Option CFT
<±7 µW
<114 µW
Keysight N8481H with Option CFT
<±0.7 µW
<11.4 µW
Keysight N8482H with Option CFT
<±0.7 µW
<11.4 µW
Table 1-4
Power sensor specifications (continued)
Power sensor
Zero drift
[a]
Measurement
[b]
noise
Summary of Contents for N1913
Page 1: ...Keysight N1913 1914A EPM Series Power Meters Service Guide...
Page 16: ...16 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 18: ...18 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 20: ...20 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...