6
Repair Guide
112
Keysight N1913/1914A Service Guide
Remove the main board (
). Use the T20 Torx
screwdriver bit to remove the five screws attaching the main
board to the bottom clamshell. Use the T20 Torx screwdriver
bit to remove the screw attaching the earth wires to the line
module. Remove the earth wires and washers, taking note of
the assembly order. Lift and remove the main board.
Figure 6-12
Remove the main board
Remove the rear panel
). Use the N1912-61807
special tooling kit to remove the fasteners on the trigger
connectors. Use the 9/32” socket to remove the GPIB
standoffs. Carefully pull the rear panel away from the main
board.
Figure 6-13
Remove the rear panel