Repair Guide
6
Keysight N1913/1914A Service Guide
113
Reassembly Instructions
Instructions
Visual
The reassembly process is simply the reverse of the
disassembly process. However, there are various points to be
aware of.
– USB/LAN connectors must rest on top of the rear panels'
EMC spring fingers.
– The position of the cable clamp depends on whether Option
101 or 003 is fitted.
– The main board connector from the PSU must be pushed
firmly to fully engage it.
– Take care not to trap any cables when fitting the top
clamshell.
– Analog recorder output connections (
). Ensure
that recorder 1 is plugged into the rear connector. Where
applicable, recorder 2 is plugged into the connector nearer
to the front.
Figure 6-14
Analog recorder output
connections
Sensor flex connections (
– A - Front, Channel A (for Option 101)
– B - Front, Channel B (for Option 101)
– C - Rear, Channel A (for Option 003)
– D - Rear, Channel B (for Option 003)
1
J28 - front Ch 1
2
J127 - front Ch 2
3
J23 - rear Ch 2
4
J18 - rear Ch 1
1
3
4
2
Summary of Contents for N1913
Page 1: ...Keysight N1913 1914A EPM Series Power Meters Service Guide...
Page 16: ...16 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 18: ...18 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 20: ...20 Keysight N1913 1914A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...