Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure
53
AC Current Gain and Offset Calibration Procedure
Review
on page 16 and
before beginning this procedure.
1
Configure each function and range shown in the table below.
2
Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see
Enter the actual applied input amplitude in Amps rms (see
3
Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the above
table. Store the new calibration constants (see
NOTE
Before the ACI Offset on a given range can be performed, the ACI gain on that
range must be performed (for example, you must perform step 35 before step
39).
SFP Cal step
Step name
Calibrator input
freq (Hz)
Calibrator input
amplitude, function
35
ACI Gain .002 A
1000
0.002 IAC
36
ACI Gain .02 A
1000
0.02 IAC
37
ACI Gain .2 A
1000
0.2 IAC
38
ACI Gain 2 A
1000
2 IAC
39
ACI Offset .002 A
1000
0.0001 IAC
40
ACI Offset .02 A
1000
0.001 IAC
41
ACI Offset .2 A
1000
0.01 IAC
42
ACI Offset 2 A
1000
0.1 IAC