Test 3: Ohms Gain Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure
25
Calibrator
input
resistance (
W
)
DMM function
DMM range (
W
)
Quick check
Error from nominal (±
W
)
24 hours
90 days
1 year
000.0000E+00
4W
W
20.00
[a]
800E-06
800E-06
1.0E-03
19.00
4W
W
20.00
Q
1.56E-03
2.51E-03
3.66E-03
000.0000E+00
4W
W
200.0000E+00
4.0E-03
4.0E-03
6.0E-03
190.0000E+00
4W
W
200.0000E+00
11.6E-03
23.0E-03
30.7E-03
000.0000E+00
4W
W
2.03
40E-03
40E-03
40E-03
1.03
4W
W
2.03
Q
97E-03
192E-03
268E-03
000.0000E+00
4W
W
20.03
400E-03
400E-03
400E-03
19.03
4W
W
20.03
970E-03
1.92E+00
2.68E+00
000.0000E+00
4W
W
200.0000E+03
4.0E+00
4.0E+00
4.0E+00
190.0000E+03
4W
W
200.0000E+03
15.4E+00
23.0E+00
36.4E+00
000.0000E+00
4W
W
2.06
Q
40E+00
60E+00
60E+00
190.0000E+03
4W
W
2.06
382E+00
630E+00
820E+00
000.0000E+00
4W
W
20.06
400E+00
600E+00
600E+00
19.06
4W
W
20.06
23.20E+03
25.30E+03
38.60E+03
000.0000E+00
4W
W
200.0000E+06
26.0E+03
30.0E+03
50.0E+03
100.0000E+06
4W
W
200.0000E+06
826.0E+03
1.0300E+06
1.3500E+06
[a] The 20
Ω
and 200 M
Ω
ranges are available only on the M9183A.