![Keysight M9181A Calibration And Verification Procedure Download Page 47](http://html.mh-extra.com/html/keysight/m9181a/m9181a_calibration-and-verification-procedure_1961485047.webp)
Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure
47
Calibration of ACV Flatness
Review
on page 16 and
before beginning this procedure.
Configuration: AC Flatness
1
Configure each function and range shown in the table below.
2
Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see
The next calibration steps perform flatness calibration at 50 kHz on each of the
ranges in the ACV. Turn ACV fast off by selecting slow filtering using the SFP.
Perform the following steps using the “calibration process” described.
3
Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the above
table.
NOTE
Before calibrating the flatness on a specific range, calibrate Gain on that range.
(For example step 17 can be performed only after step 12 is performed).
SFP Cal step
Step name
Calibrator input
freq (Hz)
Calibrator input
amplitude, function
17
ACV FComp .2 V
50,000
0.2 VAC
18
ACV FComp 2 V
50,000
2 VAC
19
ACV FComp 20 V
50,000
20 VAC
20
ACV FComp 200/300 V
50,000
160 VAC