Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure
39
The calibration count increments up to a maximum of 2
32
(over 4 billion) after
which it rolls over to 0. The value increments by one for each time the calibration
points are store to non volatile memory. Therefore, a complete calibration may
increase the value by many counts.
You can see the calibration count from the Tools menu in the SFP. To do this
remotely, use the method AGM918X_ATTR_CALIBRATE_COUNT.
Calibration Process
The following general procedure is the recommended method to complete a full
instrument calibration. The initial verification tests may be skipped if initial data is
not required.
1
2
Perform the verification tests to characterize the instrument.
3
Unsecure the instrument for calibration (
4
Perform the adjustment procedures.
5
Store the calibration constants in the calibration memory.
6
Secure the instrument’s calibration memory to prevent accidental or
unauthorized adjustments.
7
Document the new security code and calibration count in the instrument’s
maintenance records.
All adjustments are made using the SFP or the remote interface. Commands used
to perform the adjustments are listed in Calibration section of the programmer’s
guide. In addition, the SFP’s call monitor log allows you to see each calibration
message sent to the DMM.