Test 4: 2-Wire Ohms Function
Keysight M9182A/M9183A Calibration and Verification Procedure
27
Calibratior
input
resistance (
W
)
DMM function
DMM
range (
W
)
Quick
check
Error from nominal (±
W
)
24 hours
90 days
1 year
19.00
2W
W
20.00
[a]
2.56E-03
3.51E-03
4.66E-03
190.0000E+00
2W
W
200.0000E+00
12.6E-03
24.0E-03
31.7E-03
1.03
2W
W
2.03
98E-03
193E-03
269E-03
19.03
2W
W
20.03
Q
971E-03
1.921E+00
2.681E+00
190.0000E+03
2W
W
200.0000E+03
15.4E+00
23.0E+00
36.4E+00
1.06
2W
W
2.06
382E+00
630E+00
820E+00
19.06
2W
W
20.06
23.20E+03
25.30E+03
38.60E+03
100.0000E+06
2W
W
200.0000E+06
826.0E+03
1.003E+06
1.350E+06
[a] The 20
Ω
and 200 M
Ω