Oscilloscope Utility Settings
6
Keysight 1000B Series Oscilloscopes User’s Guide
131
2
Continue pressing the
Output
softkey to select the desired output condition:
The output condition can be used:
• To stop a running mask test.
• As a source for the waveform recording function (see
• As a signal on the oscilloscope back panel’s isolated
Mask Pass/Fail out
BNC.
The Mask Pass/Fail out circuit uses optical isolation. An external circuit is
required to use the signal. Before connecting to an external circuit, make sure
the maximum voltage/current does not exceed 400 V/100 mA. The output
device has no polarity limit and can be connected arbitrarily.
Fail
A mask failure sets the output.
Fail +
A mask failure sets the output and causes a beep.
Pass
A passing waveform sets the output.
Pass +
A passing waveform sets the output and causes a beep.
Figure 55
Schematic Diagram of Mask Pass/Fail out
Optically Isolated
Recommended Custom Circuit
+5.0 V
2k
Ω
BNC
connector
BNC
output
socket
PhotoMOS Relay
to internal circuit
I < ±100 mA
V < ±400 V