184 Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Standard Operation Status Group
The Operation Status Group is used to determine the specific event that set bit
7 in the Status Byte Register. This group consists of the Standard Operation
Condition Register, the Standard Operation Transition Filters (negative and
positive), the Standard Operation Event Register, and the Standard Operation
Some of the bits in this status group do not apply to all models, and return
zero when queried. See
for more information.
The SCPI command
:STAT:OPER:SUPP
can suppress the managing of this
status group and save 50 us from the switching time. Refer to the SCPI Command
Reference.
NOTE:
Some
signal
generator
models
may use
only a
subset of
the status
registers
shown.