186 Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Standard Operation Event Register
The Standard Operation Event Register latches transition events from the
condition register as specified by the transition filters. Event registers are
destructive read only. Reading data from an event register clears the content of
that register.
Standard Operation Event Enable Register
The Standard Operation Event Enable Register lets you choose which bits in
the Standard Operation Event Register set the summary bit (bit 7 of the Status
Byte Register) to 1.
Query:
STATus:OPERation[:EVENt]?
Command:
STATus:OPERation:ENABle <value>
, where
<value>
is the sum of the decimal values of the bits you want to enable.
Example:
To enable bit 5and bit 3 so that whenever either of those bits are set to 1, the Standard Operation
Status summary bit of the Status Byte Register is set to 1. Send the command
STAT:OPER:ENAB
40
(32 + 8).
Query:
STATus:OPERation:ENABle?
Response:
Decimal value of the sum of the bits previously enabled with the
STATus:OPERation:ENABle
<value>
command.