Step 5: Verify Operation of the M9381A and M9393A
Software Installation
Keysight Y1299A-004 RF PA/FEM Reference Solution Startup Guide
35
Module
Green
Orange
Red
Off
module.
The module has
returned to idle
from mod-
ulation.
load
detected
supply failure is resolved.
M9214A
The Soft Front Panel
has initialized the
connection to the
module
Missing 100 MHz
Reference.
n/a
Not connected by the SFP.
Failure in the power supplies.
Module hardware health can't
be determined until the power
supply failure is resolved.
M9308A
The Soft Front Panel
has initialized the
connection to the
module
Tuning is in progress,
or the M9308A is
unlocked from the ref-
erence
.
n/a
Not connected by the SFP.
Failure in the power supplies.
Module hardware health can't
be determined until the power
supply failure is resolved.
M9365A
The Soft Front Panel
has initialized the
connection to the
module
Missing 100 MHz
Reference.
n/a
Not connected by the SFP.
Failure in the power supplies.
Module hardware health can't
be determined until the power
supply failure is resolved.
4. Conduct a Self Test on all modules (Utilities > Self Test... > Run Self Test).
5. If the Self Test passes (see results below), go to