5
PNA-L (N5230A)
Additional options
–
Time domain (Option 010)
This option enables the PNA Series to view reflection and
transmission responses in both time or distance domain.
Use time domain to tune filters, gate out the response of
fixtures and cables, characterize the impedance of trans-
mission line, and more.
–
Frequency offset (Option 080)
This option enables the PNA Series to set the source fre-
quency independently from where the receivers are tuned.
This ability is important for two general classes of devices:
mixers (and converters) and amplifiers.
–
Scalar-calibrated converter measurements (Option 082)
With a simple setup and calibration, this application
provides the highest accuracy for conversion-loss (or gain)
measurements by combining one-port and power-meter
calibrations to remove mismatch errors. Option 080
required.
–
4-port measurement application (Option 550)
Enables full 4-port error correction and differential mea-
surements on a 2-port network analyzer with configurable
test set (Option x25). External test set required. User
installable.
–
N-port capabilities (Option 551)
Adds full N-port error correction and measurement capa-
bilities to any PNA-L network analyzer with configurable
test set (Option xx5 or x46). External test set required.
User installable.
Certification options
–
Commercial calibration certification with test data (Option UK6)
Complete set of measurements which tests unit to manufac-
turer’s published specifications. Includes calibration label,
calibration certificate, and data report. Conforms to ISO
9001.
–
ISO 17025 compliant calibration (Option 1A7)
Complete set of measurements which tests unit to manufac-
turer’s published specifications. Includes calibration label,
ISO 17025 calibration certificate, and data report, mea-
surement uncertainties and guardbands on all customer
specifications. Conforms to ISO 17025 and ISO 9001.
–
ANSI Z540 compliant calibration (Option A6J)
Complete set of measurements which tests unit to man-
ufacturer’s published specifications. Includes pre- and
post-adjustment data with measurement uncertainty
information compliant to the ANSI/NCSL Z540 standard.