Keysight PD1000A Test Fixture Operating Guide
Test Fixture Calibration
6
Click
Next
. This opens the following screen.
7
Specify the DUT connector type for both ports to ‘
PIN
’.
8
Specify the Cal Kit files. Select
TO_Fixture_In.xkt
for the Input Port 1 and
TO_Fixture_Out.xkt
for the Output Port 2. These Cal Kit files are used for all three
test fixtures (TO-220/TO-247 and SMD).
9
Click
Next
. This opens the following screen:
10
Insert the Port 1
OPEN
Cal Standard into the test fixture. The
OPEN
Cal
Standard goes between the Gate and Drain/Collector. Click
Measure
.
The remainder of this procedure involves inserting the Calibration
Standards into the Test Fixtures. Refer to
into the PD1000A TO-220/TO-247 Test Fixtures”
“Inserting the Cal Standards into the PD1000A SMD Test Fixture”
on page 29 for detailed instructions on inserting the Calibration
Standards.