22
Keysight PD1000A Test Fixture Operating Guide
Test Fixture Calibration
Install the Calibration Kit Files
The two Calibration Kit files (
TO_Fixture_In.xkt
for the Input Port and the
TO_Fixture_*.xkt
for the Output Port) are available from
www.keysight.com/find/PD1000A
. If you have not already done so, download
these files directly to the E5080A hard drive (if it has an Internet connection) or
onto a USB memory stick.
1
If you have not already done so, download the two Calibration Kit files
(
TO_Fixture_In.xkt
for the Input Port and the
TO_Fixture_*.xkt
www.keysight.com/find/PD1000A
.
–
If your E5080A has an Internet connection, download the files directly to
the E5080A hard drive. Otherwise, download the files and copy them to a
USB memory stick.
2
Press the
CAL
button.
3
Touch
CAL
Sets & Cal Kits > Cal Kit... .
4
Click the
Import
button.
5
Navigate to where you stored the two Cal Kit files.
6
Select the
TO_Fixture_In.xkt
file.
7
Click
Open
.
8
Select the ENA folder:
D:/User > PNACalKits > user
.
9
Repeat steps 3 through 6 for the
TO_Fixture_Out.xkt
file.
10
If you have not already saved the .xkt files, select both files and click
Save
to
save them to a location on the ENA hard drive. (The files, which are stored in
the ENA’s volatile memory, are otherwise lost when you cycle the power.)
Perform the Calibration
1
Connect the text fixture to the E5080A ENA as shown below. Use two
PD1000A-BST Bias Tees. See Figure 4 on page 14.
2
Setup the analyzer as follows:
Parameter
Value
Start Frequency
Set the desired Start and Stop frequencies for the
frequency range you want to measure. In general, they
should be in the range of 10 kHz to 1 GHz (or 3 GHz).
Stop Frequency
Number of Points
25
Time Domain Mode
Low-pass Step
Calibration Type
2-port SOLT