Keysight PD1000A Test Fixture Operating Guide
37
Troubleshooting
Troubleshooting
This section provides a simple functional verification procedure to test the
PD1000A Test Fixtures, cables, and Calibration Standards. This test procedures
do not verify the resistance accuracy of the Calibration Standards.
SMA Cables
Do not excessively bend, flex, or stretch the SMA cables. Check the SMA cables
regularly, and replace them if they are damaged in any way. It is possible for the
outside shielding of the cables to break.
Measuring the DC Resistance of the Calibration Standards
A simple verification test for the Calibration Standards is to measure the DC
resistance of the standards. To the TO-220 and TO-247.
Measure the resistance between the two Cal. Standard pins or SMD pads. As
shown below, the Calibration Standards should measure open (infinite
resistance), short (0
Ω
), 50
Ω
, and pass-through (Gate to Drain/Collector).
Figure 25
Damaged SMA Cable
Damaged shield
on SMA cable
Good SMA cable
Figure 26
Measuring the DC Resistance of the TO-220/TO-247 Cal Standards
Measures open
(infinite)
resistance
between pins
Measures short
(0
Ω
) resistance
between pins
Measures 50
Ω
resistance
between pins
Measures short
(0
Ω
) resistance
between pins
50
Ω
SMD
resistor
0
Ω
SMD
resistor