Keysight PD1000A Test Fixture Operating Guide
23
Test Fixture Calibration
3
Once all parameters have been set in the ENA, the calibration can be
performed. Press the
Cal
button
.
This opens the
Select Ports
screen.
4
Touch
Other Cals > Smart Cal.
5
Select
2PortCal
. Specify
Port 1
as the First Port and
Port 2
as the Second Port.
Figure 12
Calibration Setup for the PD1000A SMD Test Fixture.
Setup for the TO-220/TO-247 Fixtures is similar.
E5080A ENA
Bias T
Bias T
Network Analyzer
B2902A Precision
Source/Measure Unit
Device Test Fixture (Surface Mount Device fixture shown)
for Gate
for Drain/Collector
Summary of Contents for PD1000A
Page 2: ......
Page 6: ...vi ...
Page 8: ...viii Keysight PD1000A Test Fixture Operating Guide ...
Page 43: ......