Keysight PD1000A Test Fixture Operating Guide
25
Test Fixture Calibration
11
Insert the Port 1
SHORT
Cal Standard into the test fixture. The
SHORT
Cal
Standard goes between the Gate and Drain/Collector. Click
Measure
.
12
Insert the Port 1
50
Ω
LOAD
Cal Standard into the test fixture. The
50
Ω
LOAD
Cal Standard goes between the Gate and Drain/Collector. Click
Measure
.
13
Insert the Port 2
OPEN
Cal Standard into the test fixture. The Port 2
OPEN
Cal
Standard goes between the Source/Emitter and Drain/Collector. Click
Measure
. Note: for the TO-220/TO-247 Test Fixtures, use the same
OPEN
Cal.
Standard as before, but turn it around to fit into the correct sockets on the
test fixture.
14
Insert the Port 2
SHORT
Cal Standard into the test fixture. The
SHORT
Cal
Standard goes between the Source/Emitter and Drain/Collector. Click
Measure
. Note: for the TO-220/TO-247 Test Fixtures, use the same
SHORT
Cal.
Standard as before, but turn it around to fit into the correct sockets on the
test fixture.