Keysight NFA X-Series Getting Started and Troubleshooting Guide 91
Troubleshooting
Check the Basics
— If the previously performed alignments did not resolve the problem, press
System
,
Alignments
,
Restore Align Defaults
. Then press
System
,
Alignments
,
Align Now
All
.
— If the analyzer exhibits large amplitude errors (> 10 dB) especially at
frequencies above 10 GHz, the RF preselector might not be properly
centered. Press
Peak Search
,
AMPTD Signal Path
,
Presel Center
. If the
signal amplitude error is corrected, the preselector characterization should
be performed. Press
System
,
Alignments
,
Ad vanced
,
Characterize
Preselector
. The characterization will take several minutes and the analyzer
must not be interrupted during this time. If the analyzer is interrupted
during the characterization process, the characterization data will be
destroyed and it will be necessary to perform the entire process again.
— Is the analyzer displaying an error message? If so, refer to the Instrument
Messages Guide.
— Check if the external frequency reference is selected but not available.
Verify that it is selected by pressing
Input/Output
,
Freq Ref In
. If
External
is
selected, changing the setting to
Sense
allows the analyzer to sense the
presence of an external reference and use it only if it is available. The
frequency of the reference should be set correctly.
— If you are using a Windows program other than the instrument application,
you may notice it running slow. Place the instrument application in single
sweep/measurement.
TIP
You can get automatic electronic notification of new software releases and
other product updates/information by subscribing to the
Keysight
Technologies Test & Measurement E-Mail Notification Service
for your signal
analyzer at: