Service
3
N8480 Series Power Sensors Operating and Service Guide
49
Table 3-1
Reflection coefficient for N8480 Series power sensors (25 ºC ± 10 ºC)
Sensor Model
Frequency
System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
N8481A
10 MHz to 30 MHz
±0.007
0.157
30 MHz to 50 MHz
±0.005
0.067
50 MHz to 2 GHz
±0.005
0.040
2 GHz to 12.4 GHz
±0.012
0.073
12.4 GHz to 18 GHz
±0.013
0.105
N8482A
100 kHz to 300 kHz
±0.007
0.212
300 kHz to 1 MHz
±0.005
0.077
1 MHz to 2 GHz
±0.008
0.030
2 GHz to 6 GHz
±0.009
0.035
N8485A
10 MHz to 50 MHz
±0.011
0.143
50 MHz to 100 MHz
±0.006
0.038
100 MHz to 2 GHz
±0.005
0.025
2 GHz to 12.4 GHz
±0.009
0.064
12.4 GHz to 18 GHz
±0.011
0.085
18 GHz to 26.5 GHz
±0.014
0.114
26.5 GHz to 33 GHz
±0.020
0.139
N8487A
50 MHz to 100 MHz
±0.013
0.038
100 MHz to 2 GHz
±0.011
0.026
2 GHz to 12.4 GHz
±0.011
0.049
12.4 GHz to 18 GHz
±0.014
0.072
18 GHz to 26.5 GHz
±0.022
0.099
26.5 GHz to 40 GHz
±0.039
0.130
40 GHz to 50 GHz
±0.055
0.144
Summary of Contents for N8480 Series