Specifications and Characteristics
36
N8480 Series Power Sensors Operating and Service Guide
Zero Set, Zero Drift and Measurement Noise
Table 2-6
Zero set, zero drift and measurement noise
Sensor
Model
Sensor Option
Range
Conditions
(RH)
1
Zero Set
4
Zero Drift
2, 4
Measurement
Noise
3, 4
N8481A
N8482A
N8485A
N8486AR
N8486AQ
N8487A
Standard
Upper
Up to 70%
±63 nW
< ±7 nW
< 114 nW
Lower
Up to 70%
±25 nW
< ±3 nW
< 80 nW
Option CFT
N/A
5
Up to 70%
±63 nW
< ±7 nW
< 114 nW
N8488A
Standard
Upper
Up to 70%
±63 nW
< ±7 nW
< 114 nW
Lower
Up to 70%
±25 nW
< ±3 nW
< 80 nW
Option CFT
N/A
5
N/A
5
N/A
5
N/A
5
N/A
5
N8481B
N8482B
Standard
Upper
Up to 70%
±63
µ
W
< ±7 µW
< 114 µW
Lower
Up to 70%
±25 µW
< ±3 µW
< 80 µW
Option CFT
N/A
5
Up to 70%
±63 µW
< ±7 µW
< 114 µW
N8481H
N8482H
Standard
Upper
Up to 70%
±6.3 µW
< ±0.7 µW
< 11.4 µW
Lower
Up to 70%
±2.5 µW
< ±0.3 µW
< 8 µW
Option CFT
N/A
5
Up to 70%
±6.3 µW
< ±0.7 µW
< 11.4 µW
1 RH is the abbreviation for relative humidity.
2 Average hourly drift, at constant temperature. It is recommended to wait for 3 minutes and perform zeroing before taking
the measurement. For N8487A with serial number MY48700101 to MY54089999 and N8488A with serial number
MY50230001 to MY540899999, it is recommended to wait for 30 minutes and perform zeroing before taking
the measurement.
3 The number of averages at 16 for normal mode and 32 for x2 mode, at a constant temperature, measured over one minute
interval and two standard deviations.
4 The zero set, zero drift, and measurement noise specifications are tested at 50 MHz.
5 N/A is the abbreviation for “not applicable”.