Specifications and Characteristics
2
N8480 Series Power Sensors Operating and Service Guide
43
Supplemental Characteristics
Supplemental characteristics are intended to provide additional
information, useful in applying the power sensor by giving a typical but
not warranted performance parameters.
Figure 2-16
Typical cal factor and SWR vs. frequency
Table 2-10
Calibration factor uncertainty at
0 ºC to 55 ºC (continued)
Frequency
25 ºC ± 3 ºC
N8488A
10 MHz to 50 MHz
3.69
50 MHz to 100 MHz
2.63
100 MHz to 2 GHz
2.40
2 GHz to 12.4 GHz
2.61
12.4 GHz to 18 GHz
3.15
18 GHz to 26.5 GHz
4.50
16.5 GHz to 50 GHz
5.67
50 GHz to 67 GHz
7.18
67 GHz to 70 GHz
8.69
92
0.9
0.95
1
1.05
1.1
1.15
Typical CF
Typical SWR
Cal
Fact
or
SWR
Typical CF and SWR vs. Frequency
91
0.05
1
12
8
4
15
19
93
94
95
96
97
98
99
100
1
1.2
1.25
Frequency (GHz)
23
27
Summary of Contents for N8480 Series
Page 1: ...Keysight N8480 Series Power Sensors Operatingand Service Guide ...
Page 2: ......
Page 3: ......