Specifications and Characteristics
2
N8480 Series Power Sensors Operating and Service Guide
21
•
The second group of characteristic types describes 'statistically' the
aggregate performance of the population of products.
These characteristics describe the expected behavior of the population
of products. They do not guarantee the performance of any individual
product. No measurement uncertainty value is accounted for in the
specification. These specifications are referred to as
typical
.
•
Typical plots
are derived from a population mean from production
testing. Plot shown may vary from unit to unit and is not warranted.
For warranted specifications, refer to the individual tables.
Conditions
The power meter and power sensor meet its specifications when:
•
Stored for a minimum of two hours at a stable temperature within the
operating temperature range, and turned on for at least 30 minutes.
•
The power meter and power sensor are within their recommended
calibration periods.
•
Used in accordance to the information provided in the power meter’s
User's Guide
.