N8480 Series Power Sensors Operating and Service Guide
Cleaning
Use a clean, damp cloth to clean the body of the N8480 Series power
sensors.
Connector Care
A solution of pure isopropyl or ethyl alchohol can be used to clean
connectors but make sure to keep in mind on its flammable nature.
Clean the connector only at a static free workstation. Electrostatic
discharge to the center pin of the connector will render the power
sensor inoperative.
Clean the connector face by first using a blast of compressed air. If the
compressed air fails to remove contaminants, use a cotton swab dipped in
isopropyl or ethyl alcohol. If the swab is too big, use a round wooden
toothpick wrapped in a lint free cloth dipped in isopropyl or ethyl alcohol.
C A U T I O N
•
The RF connector bead deteriorates when contacted with any chlorinated or
aromatic hydrocarbon such as acetone, thrichlorethane, carbon tetrachloride,
and benzene.
•
Do not attempt to clean connectors with anything metallic such as pins or
paper clips.